Alcomaster PB100 Breathalyzer

The Alcomaster PB100 breathalyser uses Fuel Cell technology to provide an accurate test result. The Fuel Cell responds directly and proportionally to Blood Alcohol Concentration. The PB100 breath tester provides exceptional accuracy and a long stable lifespan. Particularly sensitive to alcohol, the fuel cell will not respond to acetone or any other breath substances.

The PB100 will automatically store 2000 tests. The data can be downloaded to a computer via the extensive yet easy to use software through an RS232 connection.

The PB100 breath analyser is fully automatic. A microprocessor controls the analysis of a fuel cell sensor, which has a high specificity to alcohol. The fuel cell has a response which is linear and directly proportional to Breath Alcohol Concentration (BAC). An accurate "base line" is established and the fuel cell temperature is monitored prior to each breath test, contributing to an accurate and precise analytical "read out".

 

 

 

 

 

 

 

ALCOMASTER PB100 TECHNICAL DATA  SHEET 

 

Sensor

High quality fuel Cell

Working temperature

0~45 Celsius

Storage temperature

0~45 Celsius

Storage humidity

Not greater than 90%

Size

165mm X 75mm X 35mm

Weight

240gm without batteries

Battery life

Not less than 500 tests

Battery

AA alkaline battery or AA Ni-Mh rechargeable battery

Display LCD size

50mm X 17mm, 128pix X 32pix

Response Time

20 seconds

Sampling system

Automatically take a deep lung sample or manually test

Optional power supply

12 volt Car Adapter or AC 12 volt Adapter

Clock

Real time clock

Units of measure

BrAC: mg/L, or any other units

Detection Range

BAC: grm%, mg/mL, mg/100ml, or any other units

Pump

Automatically actuated electronic pump system. No cocking required

Self Diagnostics

Programmed self-check assures unit is operational upon power-up

Accuracy

0.020 mg/L for all mass concentrations less than 0.400 mg/L. 5% of the measured concentration for all mass

 

concentrations greater than or equal to 0.400 mg/L and less than or equal or equal to 2.000 mg/L.

 

20% of the measured concentration for all mass concentrations greater than 2.000 mg/L

Detection Range

0.000mg/L to 3.000 mg/L